  this ndir gas analyzer features high accuracy, multiple functions and easy operation through use of a microproces- sor. it also utilizes a mass flow detector noted for its high sensitivity and reliability. this analyzer is used not only for measuring environmen- tal pollution but is also used for various processes and/or experiments. this product, combined with sampling system (zsp), has passed the type-approval test based on weight and mea- sure act. (1) use of a microprocessor provides high accuracy, mul- tiple functions and easy operation. ? maximum of 3 gas components can be calibrated with the built-in automatic calibrator (option). using a signal from a zirconiz o 2 sensor (zfk3) or from any other o 2 sensor, the gas analyzer is able to out- put a value converted into o 2 (option). alarm function to emit high/low limit contact output is also available (option). measuring range can be selected by using external single (option). zero and span can be calibrated with high accuracy, simply by pressing calibration keys. ? self-diagnosis function is included. (2) this analyzer utilizes mass flow detector featuring high sensitivity and reliability. it has 2 standard measuring ranges with a range ratio of up to 1:20. (3) three different gas components can be measured si- multaneously by connecting a zirconia o 2 sensor (zfk3), in addition to one-and two-component type sensors. (4) besides the standard measurement type, a sample swi- tching type and a differential flow type are also available. general items power supply: 100v 10% ac, 50/60hz 115, 220v ac also available on request note: refer to "code symbols" power consumption: 125va max. ambient temperature: ? to +45 c ambient humidity: 90% rh or less enclosure: steel casing, indoor-use dimensions (h x w x d): 835 x 220 x 232 mm mass{weight}: approx. 24kg finish color: munsell 2.5y 8.4/1.2 indication: 4-digit led for concentration 4-digit led for sub-indication output hold: output hold prior to manual/auto calibra- tion is possible. hold on/off is select- able. standard requirements for sample gas: temperature 0 to 50 c (dehumidification 2 c satu- ration or less) dust 0.3 m or less pressure 9.8kpa{0.1kgf/cm 2 } or less (flow rate 0.5 r /min) standard control for sample gas: calibration gas dry gas interfere control gas 2 c saturation warm-up time: about 8 hours (after power on) about 4 hours for sample switching type material of gas-contacting parts: sample cell; 304 stainless steel, neoprene rubber infrared-ray transmitting window; caf 2 or sapphire internal tubing; teflon tube, silicone tube, toaron tube gas inlet/outlet, purge gas inlet size: rc 1/4 (pt 1/4 internal thread) or npt1/4 internal thread features specifications data sheet zrg ndir type infrared gas analyzer date EDS3-110D jun. 29, 2005
2 zrg purge gas flow rate: 1 0.5 r /min purging is required when measuring gas is contained in the atmosphere or the range of co 2 is 0 to less than 50ppm. in other cases, purging should be made as necessary. scope of delivery: analyzer, test report, power fuse, cloth for cleaning infrared-ray transmitting window mounting method: flush mounting (vertical mounting) installation condition: install the analyzer at a place not exposed to direct sunlight or the radiation from a high temperature object. avoid vibration, and select a clean place free from corro- sive and/or combustible gases. if install- ing outdoors, provide a suitable casing or cover to protect the analyzer from wind, rain, etc. standard type measuring system: non-dispersion infrared-ray absorption method, single light source, double beam measurable components and measuring range: standard single-component analyzer measurable component co co 2 no so 2 ch 4 carbon monoxide carbon dioxide nitrogen monoxide sulfur dioxide methane min. measuring range [ppm] 0 to 50 0 to 5 0 to 50 0 to 50 0 to 200 measuring range see page 12 standard two-component analyzer no + so 2 co + co 2 no + co nitrogen monoxide + sulfur dioxide carbon monoxide + carbon dioxide nitrogen monoxide + carbon monoxide min. measuring range [ppm] 0 to 100/0 to 100 0 to 100/0 to 100 0 to 250/0 to 250 measuring range refer to "code symbols" measuring range: refer to table. output signal: 0 to 1v dc or 4 to 20ma dc (allowable load resistance, 550 ? or less), linear repeatability: within 0.5% of full scale (* 1 ) (within 1% of full scale) linearity: within 2% of full scale zero drift: within 2% of full scale/week (* 1 ) (within 2% of full scale/day) (* 2 ) (within 3% of full scale/week) span drift: within 2% of full scale/week (* 1 ) (within 2% of full scale/day) (* 2 ) (within 3% of full scale/week) response time: max. 50 seconds (for 90% response) in- cluding gas substitution time; time differs with the length of sample cell. measured gas flow rate: 0.5 0.25 r /min note: *( 1 ) ( ): for the range less than 50ppm *( 2 ) ( ): for 50ppm range sample switching type (sample switching type gas analyzer suited for measure- ment of low concentrations or for eliminating the effects of interfering components.) measuring system: non-dispersion infrared-ray absorption method, single light source, double beam, sample switching system measurable component: co (carbon monoxide) measuring range: 1st range [ppm] 0 to 2 0 to 5 0 to 10 0 to 25 0 to 50 0 to 100 2nd range, 3rd range, 4th range (* 4 ) [ppm] select from 5, 10, 20, 25, 50, 100 max. range rate within 1:10 1st range<2nd range<3rd range<4th range output signal: 0 to 1v dc or 4 to 20ma dc, linear, step output which changes every 50 seconds repeatability: within 1% of full scale (* 3 ) (within 2% of full scale) linearity: within 2% of full scale zero drift: within 0.5% of full scale/week (* 3 ) (within 1% of full scale/week) span drift: within 1.5% of full scale/week (* 3 ) (within 2.5% of full scale/week) response time: within 120 seconds (for 90% response) (* 3 ) (within 150 seconds) measured gas flow rate: 1 0.1 r /min (reference gas) 1 0.1 r /min (sample gas) sample switching relay drive output: 3v dc, on/off signal at 50 second in- terval (for solid state relay drive) note: *( 3 ) ( ): for 0 to 2ppm range *( 4 ) ( ): sample switching type with 4-ranges is available (max. range ratio within 1:10). in 4-range analyzer, optional func- tions of remote range, range identification, alarm and exter- nal hold are not available. differential flow type (gas analyzer suited for measurement in two modes, ab- solute concentration and concentration difference) measuring system: non-dispersion infrared-ray absorption method, single light source, double beam, differential flow system measurable components and measuring range: measurable component 1st and 2nd component carbon dioxide carbon mon- oxide co 2 co 1st range [ppm] ?50 to +50 ?00 to +100 0 to 50 0 to 100 0 to 200 0 to 250 note: there are restrictions on the reference gas conditions. output signal: 1st range ?1 to +1v dc 0 to 1v or 4 to 20ma dc co 2 co 2nd range 0 to 1v dc 0 to 1v or 4 to 20ma dc note: linear output repeatability: within 0.5% of full scale linearity: within 2% of full scale zero drift: within 2% of full scale/week span drift: within 2% of full scale/week response time: max. 50 seconds (for 90% response) in- cluding gas substitution time measured gas flow rate: 0.5 0.25 r /min (reference gas) 0.5 0.25 r /min (sample gas) 2nd range [ppm] 0 to 500 0 to 1000 none, 0 to 100, 0 to 200, 0 to 250, 0 to 500 none, 0 to 200, 0 to 250, 0 to 500 none, 0 to 500 none, 0 to 500
3 optional specifications (there are added on request. refer to the "code symbols".) o 2 correction output: an private o 2 sensor is used for correct- ing the measured gas concentration into the value at standard o 2 concentration. for obtaining the nox exhaust standard value, nox concentration and residual oxygen concentration in exhaust gas are measured simultaneously, then it is cor- rected according to the following equa- tion. (application of this equation is man- datory for the nox exhaust standard.) c= cs where c: concentration after o 2 correc- tion cs: nox measured concentration os: o 2 measured concentration on: o 2 standard concentration block diagram (example of nox measurement) 21 ?on 21 ?os nox signal 21 ?on 21 ?os ?s the o 2 measured signal is according to the private o 2 sensor (zfk) or exter- nal o 2 meter (0 to 1v dc/0 to 25% o 2 ). o 2 output signal: 0 to 1v or 4 to 20ma dc, linear o 2 correction output signal: 0 to 1v or 4 to 20ma dc, linear output can be provided for each of 1st and 2nd components alarm output: upper limit alarm contact output, spdt (1c) contact contact capacity, 250v ac, 2a (resistive load) lower limit alarm contact output, spdt (1c) contact contact capacity, 250v ac, 2a (resistive load) remote range changeover: range is changeable via external signal. range changeover input signal: 5v dc (minimum range selection at 5v input) range identification signal output: contact output, spst (1a) contact contact capacity, 250v ac, 2a (resistive load) external hold: output hold is possible with external in- put signal. external hold signal input: 5v dc (hold at 5v input) average value output: moving average value or average value is outputted at intervals of 1-hour or 4- hour (only the 1-output type for 4-hour average value is available). output signal: 0 to 1v or 4 to 20ma dc, linear. o 2 signal o 2 corrected nox [ppm] automatic calibration: zero and span are automatically calibrated at the present cycle. calibrating gas is supplied by operating the external electromagnetic valve. calibration channels: up to 3 components can be calibrated simultaneously. zero calibration point: fixed at 0% (air point and span point can be set by zirconia o 2 meter) span calibration point: 0 to 100% of full scale calibration start: built-in timer or remote start signal output hold during calibration: possible calibration gas flow mode: (1) zero gas (2) zero gas ?span gas 1 (3) zero gas ?span gas 1 ?span gas 2 (4) zero gas ?span gas 1 ?span gas 3 (o 2 ) (5) zero gas ?span gas 1 ?span gas 2 ?span gas 3 (o 2 ) calibration gas flow time: settable from 100 to 599 seconds calibration cycle: 1 to 99 hours (1-hour units) or 1 to 7 days (1-day units) calibration failure alarm: provided when fault occurs during auto calibration. contact output: under calibration; spst (1a) contact, con- tact capacity 250v ac, 2a (resistive load) calibration failure; spst (1a) contact, con- tact capacity 250v ac, 2a (resistive load) electromagnetic valve drive; spst (1a) contact, contact capacity 250v ac, 2a (resistive load) remote start: remote start signal voltage input 5v dc (start at 100msec pulse input)
4 zrg infrared light source (single light source) motor rotary sector (one-point chopper) sector synchronous signal alarm output range identification signal calibration output sv drive signal remote range input external hold input (option) output, 0 to 1v or 4 to 20ma dc operating key indication sample gas inlet sample gas ac voltage amplifier sample gas outlet gain changeover a/d cpu indicator d/a contact input,output ram rom distribution cell (interference filter) interference compensating detector trimmer reference cell reference gas inlet reference gas outlet distribution cell (interference filter) interference compensating detector trimmer infrared light source (single light source) motor rotary sector (one-point chopper) sample gas inlet sample gas ac voltage amplifier sample gas outlet sector synchronous signal gain changeover a/d cpu indicator d/a contact input, output alarm output various average value (option) (option) (option) output, 0 to 1v or 4 to 20ma dc operating key display range identification signal contact output during calibration remote range input external hold input ram rom o 2 meter o 2 correction value reference cell (differential flow type) fundamental diagram (standard type)
5 description of sample switching system the measured gas is divided into two, and in one of the flow paths, a high performance converter is provided to con- vert carbon monoxide into carbonic acid gas. here, the carbon monoxide in the measured gas is eliminated and the gas is led into one cell of the high-sensitivity infrared ana- lyzer. in the other flow path, the gas is led directly into the other cell of the analyzer. these flow paths are changed over via changeover valves sv 1 and sv 2 every 50 seconds by means of the changeover valve drive signal transmitted from the analyzer. by carry- ing out this changeover cyclically, the output of the ana- lyzer varies with the concentration of the carbon monoxide in the measured gas. use of the obtained variation width as a measured value enables improving the s/n ratio and eliminating the effect of interfering components plus zero drift. infrared light source (single light source) motor rotary sector (one-point chopper) sector synchronous signal alarm output range identification signal calibration output sv drive signal remote range input external hold input (option) output, 0 to 1v or 4 to 20ma dc operating key sample gas inlet sample cell ac voltage amplifier sample gas outlet gain changeover a/d cpu relay drive signal indicator indication d/a contact input,output ram rom distribution cell (interference filter) interference compensating detector trimmer reference cell reference gas inlet reference gas outlet co measured gas 1ry filter pump sv 4 reference gas sv 3 sv 1 sv 2 sv 2 sv 1 t emperature controller co/co 2 converter flowrator 2ry filter zrg3 reference cell sample cell sample switching type analyzer relay output signal fundamental diagram (sample switching type)
6 zrg (standard single-component analyzer) code symbols y -- 2 y g r z0 21 14 15 16 17 18 19 20 9101 11213 2 134 5678 description measurable components so 2 co co 2 ch 4 no other non-standard items power supply piping connection 100v ac 50/60hz rc 1/4 115v ac 50/60hz npt 1/4 220v ac 50/60hz rc 1/4 220v ac 50/60hz npt 1/4 1st component, 1st range 0 to 5ppm 0 to 10ppm 0 to 15ppm 0 to 20ppm 0 to 25ppm 0 to 50ppm 0 to 100ppm 0 to 200ppm 0 to 250ppm 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 100% other non-standard items measuring method single-component analyzer single-component analyzer (approved by weight and measure act) note: (7) corrected o 2 is designated by 18th digit. (8) external o 2 analyzer is required 0 to 1v linear output. output signal 0 to 1v dc linear 4 to 20ma dc linear optional functions [1] automatic calibration _ ? _ _ ? ? _ ? average value output _ _ ? _ ? _ ? ? range identification alarm, remote range, external hold _ _ _ ? _ ? ? ? a b d e p z 1 5 w v 0 1 t a b c d e f g h j k l m n p r z 0 2 3 4 a b y a b c d y a b d f h k m note: (1) when selecting a range, be sure to refer to the table of "measuring range". (2) the specifications for dioxin co analyzer of 200ppm max. are different in some points. note: (3) details of average value output depend on 16th and 17th digits. (4) average value output includes moving average value output. optional functions [2] o 2 indication (external o 2 analyzer connection) _ ? _ ? _ o 2 indication (zfk connection) _ _ ? _ ? range identification alarm, remote range _ _ _ ? ? note: (5) in the case of o 2 indication, corrected o 2 is outputted as standard specification. (6) standard type of o 2 analyzer has 2 ranges (0 to 10/25%)
7 (standard single-component analyzer) (cont'd) y -- 2 y g r z0 21 14 15 16 17 18 19 20 9101 11213 2 134 5678 description average value output time without 1-hour moving average value output 4-hour moving average value output 1-hour average value output 4-hour average value output 1st component, 2nd range 2nd range without 0 to 10ppm 0 to 15ppm 0 to 20ppm 0 to 25ppm 0 to 50ppm 0 to 100ppm 0 to 200ppm 0 to 250ppm 0 to 500ppm 0 to 1000ppm 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% 0 to 10% 0 to 20% 0 to 50% 0 to 1 00% other non-standard items note: (15) data should be put on data code. (16) quick response type is available only when gas density is 20% or more. average value output object component 1st component instantaneous value o 2 corrected _ ? _ _ _ ? o 2 corrected standard value without 4% 5% 6% 7% 10% 11% 12% 15% other non-standard items (specify within 0 to 19%) measuring gas kind atmospheric gas combustible exhaust gas (dioxin included) converter exhaust gas other non-standard items y v 0 1 t a b c d e f g h j k l m n p r z 0 1 4 5 6 y a d z a y 4 5 6 7 a b c f z e y f y g y z z note: (9) 2nd range>1st range range rate: within 1:20 note: (11) average value output (option) designated only by 12th digit. (12) for selection of 4-hour moving average value output and 4-hour average value output, only the 1-output type is available. note: (13) for designation of "z", the data must be put on the data code. note: (14) for designation of "z", gas components specifications should be attached. note: (10) average value output (option) designated only by 12th digit. non-standard spec. other non-standard items quick response
8 zrg (standard two-component analyzer) -- 2 g r z0 21 14 15 16 17 18 19 20 9101 11213 2 134 5678 description measurable components (1st component) + (2nd component) no + so 2 co + co 2 no + co other non-standard items code b c d e f range value 0 to 100 ppm 0 to 200 ppm 0 to 250 ppm 0 to 500 ppm 0 to 1000 ppm code g h j k l range value 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% code m n p r z range value 0 to 10 % 0 to 20 % 0 to 50 % 0 to 100 % other non-standard items power supply piping connection 100v ac 50/60hz rc 1/4 115v ac 50/60hz npt 1/4 220v ac 50/60hz rc 1/4 220v ac 50/60hz npt 1/4 1st component, 1st range note: (1) for range selection, be sure to refer to the table of "measuring range". (2) no + co min. range: 0 to 250ppm code b c d e f range value 0 to 100 ppm 0 to 200 ppm 0 to 250 ppm 0 to 500 ppm 0 to 1000 ppm code g h j k l range value 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% code m n p r z range value 0 to 10 % 0 to 20 % 0 to 50 % 0 to 100 % other non-standard items 2nd component, 1st range note: (3) for range selection, be sure to refer to the table of "measuring range". (4) no + co min. range: 0 to 250ppm measuring method two-component analyzer two-component analyzer (approved by weight and measure act) note: (10) corrected o 2 is designated by 18th digit. (11) external o 2 analyzer is required 0 to 1v linear output. output signal 0 to 1v dc linear 4 to 20ma dc linear other non-standard items optional functions [1] automatic calibration _ ? _ _ ? ? _ ? note: (7) this function applies to 1st and 2nd components. average value output _ _ ? _ ? _ ? ? range identification alarm, remote range, external hold _ _ _ ? _ ? ? ? f g h z 2 6 b z 0 2 3 4 a b z y a b c d y a c e g j l n note: (5) details of average value output depend on 16th and 17th digits. (6) average value output includes moving average value output. optional functions [2] o 2 indication (external o 2 analyzer connection) _ ? _ ? _ o 2 indication (zfk connection) _ _ ? _ ? range identification alarm, remote range _ _ _ ? ? note: (8) in the case of o 2 indication, corrected o 2 is outputted as standard specification. (9) standard type of o 2 analyzer has 2 ranges (0 to 10/25%) code y c d e f range value without 0 to 200 ppm 0 to 250 ppm 0 to 500 ppm 0 to 1000 ppm code g h j k l range value 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% code m n p r z range value 0 to 10 % 0 to 20 % 0 to 50 % 0 to 100 % other non-standard items 1st component, 2nd range note: (12) 2nd range>1st range range rate: within 1:20 y z b z
9 (standard two-component analyzer) (cont'd) note: (19) data should be put on data code. (20) quick response type is available only when gas density is 20% or more. -- 2 g r z0 21 14 15 16 17 18 19 20 9101 11213 2 134 5678 description average value output time without 1-hour moving average value output 4-hour moving average value output 1-hour average value output 4-hour average value output 2nd component, 2nd range code y c d e f note: (13) 2nd range>1st range range rate: within 1:20 range value without 0 to 200 ppm 0 to 250 ppm 0 to 500 ppm 0 to 1000 ppm code g h j k l range value 0 to 2000ppm 0 to 5000ppm 0 to 1% 0 to 2% 0 to 5% code m n p r z range value 0 to 10 % 0 to 20 % 0 to 50 % 0 to 100 % other non-standard items o 2 correct standard value without 4% 5% 6% 7% 10% 11% 12% 15% other non-standard items (specify within 0 to 19%) measuring gas kind atmospheric gas combustible exhaust gas (dioxin included) converter exhaust gas other non-standard items y z 0 1 4 5 6 z a y 4 5 6 7 a b c f z y a b c d e f g h average value output object component 1st component instantaneous value _ ? ? ? _ _ _ _ _ o 2 corrected _ _ _ _ ? ? ? _ _ 2nd component instantaneous value _ _ ? _ _ ? _ ? _ note: (15) average value output (option) designated only by 12th digit. (16) for selection of 4-hour moving average value output and 4-hour average value output, only the 1-output type is available. o 2 corrected _ _ _ ? _ _ ? _ ? e y f y g y z z note: (14) average value output (option) designated only by 12th digit. note: (17) for designation of "z", the data must be put on the data code. note: (18) for designation of "z", gas components specifications should be attached. non-standard spec. other non-standard items quick response
10 zrg (sample switching type) yy y -- - 2 y g r z00 21 14 15 16 17 18 19 20 9101 11213 2 134 5678 description measurable components co power supply piping connection 100v ac 50/60hz rc 1/4 115v ac 50/60hz npt 1/4 220v ac 50/60hz rc 1/4 220v ac 50/60hz npt 1/4 1st range 0 to 2ppm 0 to 5ppm 0 to 10ppm 0 to 25ppm 0 to 50ppm 0 to 100ppm other non-standard items measuring method sample switching system output signal 0 to 1v dc linear 4 to 20ma dc linear optional functions [1] automatic calibration _ _ ? ? range identification alarm, remote range, external hold _ ? _ ? b 3 x w v t a b z 0 2 3 4 a b y w v 1 2 3 4 5 v t a 6 7 8 9 t a b 8 m n p a b n b y d a h y y y y e z y z z 2nd range/3rd range/4th range 2nd range 5 10 20 5 5 10 5 10 25 50 10 10 25 10 25 50 100 25 25 50 25 50 100 50 100 measuring gas kind atmospheric gas other non-standard items non-standard spec. other non-standard items note: (1) for designation of "z", gas components specifications should be attached. note: (2) data should be put on data code. 3rd range 10 20 20 10 25 50 50 25 50 100 100 50 100 4th range 20 50 100 [unit: ppm] for 1st range 2ppm for 1st range 5 ppm for 1st range 10ppm for 1st range 25ppm for 1st range 50ppm
11 (differential flow system) y0yy -- - 2 y g r z0 21 14 15 16 17 18 19 20 9101 11213 2 134 5678 description power supply piping connection 100v ac 50/60hz rc 1/4 115v ac 50/60hz npt 1/4 220v ac 50/60hz rc 1/4 220v ac 50/60hz npt 1/4 measuring method differential flow system note: (3) data should be put on data code. output signal 0 to 1v dc linear 4 to 20ma dc linear 2nd range [unit : ppm] without 500 . for 1st range co2 350 50 1000 . for 1st range co2 350 100 100 200 250 500 200 250 500 500 . for 1st range co 200ppm 500 . for 1st range co 250ppm measuring gas kind atmospheric gas other non-standard items 4 0 2 3 4 a b z y e f b c d e c d e e e y y d y a y h y optional functions (1) automatic calibration _ _ ? ? range identification, alarm, remote range, external hold _ ? _ ? d q d s b a b b b c b d measurable component co 2 (carbon dioxide) co (carbon monoxide) 1st range 350 50ppm 350 100ppm 0 to 50ppm 0 to 100ppm 0 to 200ppm 0 to 250ppm e y z y note: (1) voltage output signal only for 5th and 6th digits, dq or ds (1st range: ?v to +1v) (2nd range: 0 to 1v) for 1st range co 50ppm for 1st range co 100ppm note: (2) for designation of "z", gas components specifications should be attached. non-standard spec. other non-standard items
12 zrg ndir type infrared gas analyzer 5ppm ? ? ? ? ? ? ? ? 50ppm ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 15ppm ? ? ? ? ? ? ? ? 20ppm ? ? ? ? ? ? ? ? 25ppm ? ? ? ? ? ? ? ? 100ppm ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 10ppm ? ? ? ? ? ? ? ? 200ppm ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 250ppm ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 500ppm ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 0.1% ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 0.2% ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 0.5% ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 1% ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 2% ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 5% ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 10% ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 20% ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? 50% ? ? ? ? ? 100% ? ? ? 1st measuring range measuable component 2nd range x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 x 0 x 2 x 2.5 x 4 x 5 x 8 x 10 x 20 co co 2 ch 4 so 2 no remarks: (1) ? : standard measuring range (2) : consult us regarding manufacture availability, price and delivery period. (3) : not manufactured no 100 ppm ? ? ? ? so 2 100ppm 200ppm 250ppm 500ppm 200 ppm ? ? ? ? 250 ppm ? ? ? ? 500 ppm ? ? ? ? remarks : (1) ? : scope of manufacture for 1st range (2) 2nd range type is available provided the range ratio of 1st or 2nd component is within 1:20. (max. 2000ppm for no analyzer) no co 250ppm 500ppm 1000ppm 500 ppm ? ? ? 1000 ppm ? ? ? 250 ppm ? ? ? remarks : (1) 2nd range type is available provided the range ratio of 1st or 2nd component is within 1:20. (max. 2000ppm for no analyzer) co 2 100ppm 200ppm 250ppm 500ppm 0.1% 0.2% 0.5% 1% 2% 5% 10% 20% 50% 100% co 100 ppm ? ? ? ? ? ? 200 ppm ? ? ? ? ? ? 250 ppm ? ? ? ? ? ? 500 ppm ? ? ? ? ? ? ? 0.1 % ? ? ? ? ? ? ? ? 0.2 % ? ? ? ? ? ? ? ? ? 0.5 % ? ? ? ? ? ? ? ? ? ? 1 % ? ? ? ? ? ? ? ? ? 2 % ? ? ? ? ? ? ? ? ? 5 % ? ? ? ? ? ? ? ? ? ? 10 % ? ? ? ? ? ? ? ? ? 20 % ? ? ? ? ? ? ? ? 100 % ? ? ? ? ? ? ? 50 % ? ? ? ? ? ? ? remarks : (1) ? : scope of manufacture for 1st range ? : scope of manufacture for 2nd range for co and co 2 (maximum range) (2) 2nd range type is available provided the range ratio of 1st or 2nd component is within 1:20.
13 outline diagram (unit:mm) zrg reference gas inlet * purge gas inlet reference gas outlet * sample gas inlet sample gas outlet < front panel > < side panel > < top panel > external terminal a/o option 1 option 2 auto cal m3.5 screw approx.230 rc or npt1/4 note * lid is mounted when reference gas is not used. operation panel rock bracket po we r supply terminal (inside) cable inlet 22 (with rubber grommet) cable inlet 22 (with rubber grommet) 140 220 p 785 20 232 813 835 813 140 3- 12hole mounting screw m10 should be used. see from p 61.5 94 mounting size
14 zrg external connection diagram + - + - + - + - + - + - + - + - + + + - + - + - + + 0 to 1v or 4 to 20ma dc (instan- taneous value) 0 to 1v or 4 to 20ma dc (moving average value) (option) 0 to 1v or 4 to 20ma dc (instan- taneous value) 0 to 1v or 4 to 20ma dc (moving average value) (option) external hold 1st component remote range input (option) o 2 remote range input (option) 1st component range identification contact output (option) 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 1st component 2nd component 2nd component remote range input (option) r 1 r 1 2nd component range identification contact output (option) o 2 remote range input (option) a/c o 2 sensor input o 2 0 to 1v or 4 to 20ma dc (instan- taneous value) 1st component o 2 corrected value 0 to 1v dc or 4 to 20ma dc o 2 2nd component corrected value 0 to 1v dc or 4 to 20ma dc f ault option 1 option 2 auto cal 1st component upper limit alarm contact output (option) 2nd component upper limit alarm contact output (option) o 2 upper limit alarm contact output (option) 1st component lower limit alarm contact output (option) 2nd component lower limit alarm contact output (option) o 2 lower limit alarm contact output (option) calibration contact output zero gas contact output span gas-1 contact output span gas-2 contact output span gas-3 contact output auto calibration failure contact output pump on/off contact remote start input sv 2 relay drive output signal (sample switching type only) sv relay drive output signal (differential or sample switching type only) note: for relay drive output signal terminal, only the sample switching type of differential flow system is used. terminal block should be used at the position of auto-cal terminal block. ground power supply (installed in the main unit) power supply
15 private zirconia o 2 s ensor


▲Up To Search▲   

 
Price & Availability of EDS3-110D

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X